CVEs (2)

  • CVE-2024-12248CriJan 30, 2025
    risk 0.64cvss 9.8epss 0.03

    Contec Health CMS8000 Patient Monitor is vulnerable to an out-of-bounds write, which could allow an attacker to send specially formatted UDP requests in order to write arbitrary data. This could result in remote code execution.

  • CVE-2025-0683MedJan 30, 2025
    risk 0.38cvss 5.9epss 0.00

    In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.