Description
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
Patches
0No patches discovered yet.
Vulnerability mechanics
AI mechanics synthesis has not run for this CVE yet.
References
4- www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/nvd
- www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01nvd
- www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoornvd
- www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communicationnvd