Plum A\+3 Infusion System Firmware

by Pifzer

CVEs (4)

  • CVE-2015-3956CriMar 25, 2019
    risk 0.64cvss 9.8epss 0.01

    Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices…

  • CVE-2015-3954CriMar 25, 2019
    risk 0.64cvss 9.8epss 0.02

    Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the…

  • CVE-2015-3953CriMar 25, 2019
    risk 0.64cvss 9.8epss 0.02

    Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the…

  • CVE-2015-3952HigMar 25, 2019
    risk 0.49cvss 7.5epss 0.01

    Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the…