Unrated severityNVD Advisory· Published Mar 25, 2019· Updated Aug 6, 2024
CVE-2015-3953
CVE-2015-3953
Description
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Affected products
1- Range: <= 13.4
Patches
0No patches discovered yet.
Vulnerability mechanics
AI mechanics synthesis has not run for this CVE yet.
References
1- ics-cert.us-cert.gov/advisories/ICSA-15-161-01mitrex_refsource_MISC
News mentions
0No linked articles in our index yet.