Medium severity4.0NVD Advisory· Published Jun 2, 2026· Updated Jun 4, 2026
CVE-2019-25723
CVE-2019-25723
Description
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
AI Insight
LLM-synthesized narrative grounded in this CVE's description and references.
Affected products
1- Range: 2.00 - 2.02
Patches
Vulnerability mechanics
References
2News mentions
1- Dräger Medical Devices: 15 Vulnerabilities Disclosed in 24-Hour WindowVypr Intelligence · Jun 2, 2026