Bd Alaris™ Point Of Care Unit (pcu) Model 8015
CVEs (3)
| CVE | Vendor / Product | Sev | Risk | CVSS | EPSS | KEV | Published | Description |
|---|---|---|---|---|---|---|---|---|
| CVE-2023-30560 | Med | 0.44 | 6.8 | 0.00 | Jul 13, 2023 | The configuration from the PCU can be modified without authentication using physical connection to the PCU. | ||
| CVE-2023-30561 | Med | 0.40 | 6.1 | 0.00 | Jul 13, 2023 | The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running. | ||
| CVE-2023-30559 | Med | 0.34 | 5.2 | 0.00 | Jul 13, 2023 | The firmware update package for the wireless card is not properly signed and can be modified. |
- risk 0.44cvss 6.8epss 0.00
The configuration from the PCU can be modified without authentication using physical connection to the PCU.
- risk 0.40cvss 6.1epss 0.00
The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running.
- risk 0.34cvss 5.2epss 0.00
The firmware update package for the wireless card is not properly signed and can be modified.