High severity7.5NVD Advisory· Published Jan 30, 2025· Updated Apr 15, 2026
CVE-2025-0626
CVE-2025-0626
Description
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
Patches
0No patches discovered yet.
Vulnerability mechanics
AI mechanics synthesis has not run for this CVE yet.
References
4- www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/nvd
- www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01nvd
- www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoornvd
- www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communicationnvd