VYPR
Unrated severityNVD Advisory· Published Dec 14, 2020· Updated May 22, 2025

Medtronic MyCareLink Smart Improper Authentication

CVE-2020-25183

Description

Medtronic MyCareLink Smart 25000 contains

an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.

AI Insight

LLM-synthesized narrative grounded in this CVE's description and references.

Affected products

2

Patches

Vulnerability mechanics

References

2

News mentions

0