CVE-2019-6538
Description
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
AI Insight
LLM-synthesized narrative grounded in this CVE's description and references.
Affected products
47- cpe:2.3:o:medtronic:amplia_crt-d_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:carelink_2090_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:carelink_monitor_firmware:2490c:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:claria_crt-d_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:compia_crt-d_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:concerto_crt-d_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:concerto_ii_crt-d_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:consulta_crt-d_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:evera_icd_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:maximo_ii_crt-d_and_lcd_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:mirro_icd_firmware:-:*:*:*:*:*:*:*
cpe:2.3:o:medtronic:mycarelink_monitor_firmware:24950:*:*:*:*:*:*:*+ 1 more
- cpe:2.3:o:medtronic:mycarelink_monitor_firmware:24950:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:mycarelink_monitor_firmware:24952:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:nayamed_nd_icd_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:primo_icd_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:protecta_icd_and_crt-d_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:secura_icd_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:virtuoso_icd_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:virtuoso_ii_icd_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:visia_af_icd_firmware:-:*:*:*:*:*:*:*
- cpe:2.3:o:medtronic:viva_crt-d_firmware:-:*:*:*:*:*:*:*
(expand)+ 1 more
- (no CPE)
- (no CPE)range: All versions
24950, 24952+ 1 more
- (no CPE)range: 24950, 24952
- (no CPE)range: 24950
- Medtronic/Amplia CRT-Dv5Range: All versions
- Medtronic/Brava CRT-Dv5Range: All versions
- Range: 2490C
- Medtronic/Claria CRT-Dv5Range: All versions
- Medtronic/Compia CRT-Dv5Range: All versions
- Medtronic/Concerto CRT-Dv5Range: All versions
- Medtronic/Concerto II CRT-Dv5Range: All versions
- Medtronic/Conexus Radio Frequency Telemetry Protocolv5Range: All versions
- Medtronic/Consulta CRT-Dv5Range: All versions
- Medtronic/Evera ICDv5Range: All versions
- Medtronic/Maximo II CRT-Dv5Range: All versions
- Medtronic/Maximo II ICDv5Range: All versions
- Medtronic/Mirro ICDv5Range: All versions
- Medtronic/Mirro MRI ICDv5Range: All versions
- Medtronic/Nayamed ND ICDv5Range: All versions
- Medtronic/Primo ICDv5Range: All versions
- Medtronic/Protecta ICD, Protecta CRT-Dv5Range: All versions
- Medtronic/Secura ICDv5Range: All versions
- Medtronic/Virtuoso ICDv5Range: All versions
- Medtronic/Virtuoso II ICDv5Range: All versions
- Medtronic/Visia AF ICDv5Range: All versions
- Medtronic/Viva CRT-Dv5Range: All versions
Patches
Vulnerability mechanics
References
2- www.securityfocus.com/bid/107544nvdThird Party AdvisoryVDB Entry
- ics-cert.us-cert.gov/advisories/ICSMA-19-080-01nvdMitigationThird Party AdvisoryUS Government Resource
News mentions
0No linked articles in our index yet.