Critical severity9.8NVD Advisory· Published Mar 26, 2019· Updated Jun 17, 2026
CVE-2014-5433
CVE-2014-5433
Description
An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
AI Insight
LLM-synthesized narrative grounded in this CVE's description and references.
Affected products
2=6.05+ 1 more
- (no CPE)range: =6.05
- (no CPE)range: 6.05 (model 35700BAX) with wireless battery module (WBM) version 16
Patches
Vulnerability mechanics
References
1- ics-cert.us-cert.gov/advisories/ICSA-15-181-01nvdMitigationThird Party AdvisoryUS Government Resource