CVE-2014-5431
Description
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.
AI Insight
LLM-synthesized narrative grounded in this CVE's description and references.
Affected products
26.05+ 1 more
- (no CPE)range: 6.05
- (no CPE)range: 6.05 (model 35700BAX) with wireless battery module (WBM) version 16
Patches
Vulnerability mechanics
References
1- ics-cert.us-cert.gov/advisories/ICSA-15-181-01nvdMitigationThird Party AdvisoryUS Government Resource
News mentions
0No linked articles in our index yet.